SU-FF-I-40: Proton Computed Tomography Reconstruction Using Compressed Sensing and Prior Image Constrained Compressed Sensing
- Source :
- Medical Physics. 36:2443-2443
- Publication Year :
- 2009
- Publisher :
- Wiley, 2009.
-
Abstract
- Purpose: To study the feasibility of protoncomputed tomography(CT)reconstruction using compressed sensing (CS) and prior image constrained compressed sensing (PICCS). ProtonCTimages can be used for pre‐therapy planning, image guidance and registration verification. Method and Materials: Projections of 200 MeV proton beams onto an ellipsoid phantom was simulated using Geant4 Monte Carlo simulation toolkit. The position and energy of the entrance and exit protons were recorded. Straight‐line path (SLP) estimation was used to represent proton paths, and simultaneous algebraic reconstruction technique (SART) with CS was used to reconstruct a protonstopping powerimage for a 2‐mm thick slice of the phantom. PICCS was used to reconstruct the image from highly undersampled data with an accurate and well‐registered prior image. A Gradient transform was used to yield a sparse data set for CS and PICCS. Results: SART with CS reconstructed a 320×320 protonstopping powerimage of the central slice of the phantom after 10 iterations. A proton/pixel ratio of 0.2 is sufficient to reconstruct an image of correct geometry. The average protonstopping power of the reconstructed materials cortical bone, water, and air were found to agree with the expected values from ICRU Report 49 within 8.3%, 0.6%, and 3.8% respectively. Employing a prior image and PICCS in the reconstruction, a proton/pixel ratio as low as 0.05 was found to be sufficient, and the reconstruction time of less than 2 minutes was achieved using a serial algorithm. Reconstruction artifacts in the images were minimal. Conclusion: With CS, or with PICCS plus a prior image, SART can reconstruct a protonCTimage of good quality within minutes. This paves the road to a clinically feasible approach toward low‐dose pre‐proton therapy treatment planning and image guidance using a fast‐reconstructed protonCTimage with a well‐registered kV‐CT prior image.
- Subjects :
- Proton computed tomography
medicine.diagnostic_test
Computer science
business.industry
Image registration
Computed tomography
General Medicine
Iterative reconstruction
Imaging phantom
Compressed sensing
medicine.anatomical_structure
Simultaneous Algebraic Reconstruction Technique
Pixel aspect ratio
Medical imaging
medicine
Cortical bone
Computer vision
Tomography
Artificial intelligence
Radiation treatment planning
business
Nuclear medicine
Subjects
Details
- ISSN :
- 00942405
- Volume :
- 36
- Database :
- OpenAIRE
- Journal :
- Medical Physics
- Accession number :
- edsair.doi...........e3d9d55d0b03dd6e02b34df366d4704a
- Full Text :
- https://doi.org/10.1118/1.3181159