Back to Search
Start Over
JCPDS—International Centre for Diffraction Data round robin study of silver behenate. A possible low-angle X-ray diffraction calibration standard
- Source :
- Powder Diffraction. 10:91-95
- Publication Year :
- 1995
- Publisher :
- Cambridge University Press (CUP), 1995.
-
Abstract
- A task group of the JCPDS—International Center for Diffraction Data (ICDD) was established with the charge of investigating the use of silver behenate, CH3(CH2)20COO·Ag, as a possible low-angle calibration standard for powder diffraction applications. Utilizing several data collection and analysis techniques, long-period spacing (d001) values with a range of 58.219–58.480 Å were obtained. Using the same collected data and one data analysis refinement calculation method resulted in dm values with a range of 58.303–58.425 Å. Data collected using a silicon internal standard and the same singular data analysis calculation method provided d001 values with a range of 58.363–58.381 Å.
- Subjects :
- Diffraction
Radiation
Materials science
Silicon
business.industry
chemistry.chemical_element
Condensed Matter Physics
chemistry.chemical_compound
Optics
chemistry
International Centre for Diffraction Data
X-ray crystallography
Calibration
Range (statistics)
General Materials Science
business
Instrumentation
Powder diffraction
Silver behenate
Subjects
Details
- ISSN :
- 19457413 and 08857156
- Volume :
- 10
- Database :
- OpenAIRE
- Journal :
- Powder Diffraction
- Accession number :
- edsair.doi...........e35e09387b18e4ceb8b4a6566ce7d9ec
- Full Text :
- https://doi.org/10.1017/s0885715600014421