Back to Search
Start Over
Development of two-dimensional multiwire-type neutron detector system with individual line readout and optical signal transmission
- Source :
- Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 726:169-174
- Publication Year :
- 2013
- Publisher :
- Elsevier BV, 2013.
-
Abstract
- A multiwire-type two-dimensional neutron detector system with a sensitive area of 128×128 mm 2 is developed for use in the Materials and Life Science Experimental Facility at the Japan Proton Accelerator Research Complex. The system can achieve a short response time and high spatial resolution using the individual line readout method. Optical devices have been incorporated in the system for long-distance signal transmission and insulation between a detector head in the neutron shielding and signal processing circuits in the data acquisition room. The detector system exhibits a pulse-pair resolution of 1 μs, an average spatial resolution of less than 2 mm full width at half-maximum in the sensitive region, and a two-dimensional homogeneity of 8.3% in all pixels.
Details
- ISSN :
- 01689002
- Volume :
- 726
- Database :
- OpenAIRE
- Journal :
- Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
- Accession number :
- edsair.doi...........e150bafecbbe66d3fe31a851cec3beb6
- Full Text :
- https://doi.org/10.1016/j.nima.2013.05.118