Back to Search Start Over

Increased signal separation upgrade permits preliminary electron anisotropy measurements with Heliotron J multi-pass Thomson diagnostic

Authors :
Dechuan Qiu
Takashi Minami
Shinichiro Kado
Shigeru Inagaki
Chenyu Wang
Ryo Matsutani
Shunsuke Toba
Maoyuan Luo
Pengfei Zhang
Naoki Kenmochi
Shinsuke Ohshima
Shinji Kobayashi
Hiroyuki Okada
Tohru Mizuuchi
Shigeru Konoshima
Ryo Yasuhara
Kazunobu Nagasaki
Source :
Review of Scientific Instruments. 94:023510
Publication Year :
2023
Publisher :
AIP Publishing, 2023.

Abstract

A signal separation system is constructed on the multi-pass Thomson scattering system of Heliotron J to solve the problem of overlapping scattered light signals for the electron temperature anisotropy measurement. The phenomenon of overlapping scattered light signals is relieved by operating the signal separation system. A Raman scattering experiment is undertaken to verify the separation effect of the signal separation system. Two scattered light signals corresponding to two adjacent incidences of one laser shot were extended to 104 ns. Moreover, we applied the multi-pass Thomson scattering system with signal separation system to the electron temperature anisotropy measurement. No anisotropy was observed within the error bars in the initial experiment.

Subjects

Subjects :
Instrumentation

Details

ISSN :
10897623 and 00346748
Volume :
94
Database :
OpenAIRE
Journal :
Review of Scientific Instruments
Accession number :