Back to Search Start Over

THEORY OF KLUMB-SCHWARZ TYPE RADIOMETER GAUGE

Authors :
Q M Pan
M Y Wang
Z Y Lee
Source :
Acta Physica Sinica. 19:90
Publication Year :
1963
Publisher :
Acta Physica Sinica, Chinese Physical Society and Institute of Physics, Chinese Academy of Sciences, 1963.

Abstract

Klumb-Schwarz type radiometer gauge (Knudsen gauge) is an absolute gauge for high vacuum measurements, but since its theoretical sensitivity is so far unknown, in practice it needs calibration against another absolute gauge (e. g. McLeod). In this paper a simple theory to determine the sensitivity of a Klumb-Schwarz gauge is described. The optimum angle of the vane is decided as 24° by theoretical considerations. It is further shown that in this gauge the force due to electromagnetic radiations limits the lower range and is something in analogue with the soft X-ray effect in conventional hot-cathode ionization gauges.

Details

ISSN :
10003290
Volume :
19
Database :
OpenAIRE
Journal :
Acta Physica Sinica
Accession number :
edsair.doi...........d59a2b4fff322784c805ced050fef45a