Back to Search
Start Over
Analysis of eddy currents induced by transverse and longitudinal gradient coils in different tungsten collimators geometries for SPECT/MRI integration
- Source :
- Magnetic Resonance in Medicine. 74:1780-1789
- Publication Year :
- 2014
- Publisher :
- Wiley, 2014.
-
Abstract
- Purpose We investigated the temporal variation of the induced magnetic field due to the transverse and the longitudinal gradient coils in tungsten collimators arranged in hexagonal and pentagonal geometries with and without gaps between the collimators. Methods We modeled x-, y-, and z-gradient coils and different arrangements of single-photon emission computed tomography (SPECT) collimators using FEKO, a three-dimensional electromagnetic simulation tool. A time analysis approach was used to generate the pulsed magnetic field gradient. The approach was validated with measurements using a 7T MRI scanner. Results Simulations showed an induced magnetic field representing 4.66% and 0.87% of the applied gradient field (gradient strength = 500 mT/m) for longitudinal and transverse gradient coils, respectively. These values can be reduced by 75% by adding gaps between the collimators for the pentagonal arrangement, bringing the maximum induced magnetic field to less than 2% of the applied gradient for all of the gradient coils. Conclusion Characterization of the maximum induced magnetic field shows that by adding gaps between the collimators for an integrated SPECT/MRI system, eddy currents can be corrected by the MRI system to avoid artifact. The numerical model was validated and was proposed as a tool for studying the effect of a SPECT collimator within the MRI gradient coils. Magn Reson Med, 2014. © 2014 Wiley Periodicals, Inc.
- Subjects :
- Scanner
Materials science
FEKO
medicine.diagnostic_test
business.industry
Physics::Medical Physics
Collimator
law.invention
Magnetic field
Transverse plane
Nuclear magnetic resonance
Optics
law
Eddy current
medicine
Radiology, Nuclear Medicine and imaging
Vector field
business
Emission computed tomography
Subjects
Details
- ISSN :
- 07403194
- Volume :
- 74
- Database :
- OpenAIRE
- Journal :
- Magnetic Resonance in Medicine
- Accession number :
- edsair.doi...........d2d62213cb5ed6474fe13274323d4ecc
- Full Text :
- https://doi.org/10.1002/mrm.25534