Back to Search Start Over

Two‐dimensional, remote micropositioner for a scanning tunneling microscope

Authors :
David W. Abraham
Eric Ganz
H. Jonathon Mamin
John Clarke
Source :
Review of Scientific Instruments. 56:2168-2170
Publication Year :
1985
Publisher :
AIP Publishing, 1985.

Abstract

A simple walker is described that is suitable as the coarse positioner in a scanning tunneling microscope. The walker is fabricated from a single piece of piezoelectric ceramic and has feet coated with thin metal and insulating films. The walker can move in either of two orthogonal, horizontal directions on an insulated, metallized quartz substrate in steps that can be varied from 25 to 400 nm.

Details

ISSN :
10897623 and 00346748
Volume :
56
Database :
OpenAIRE
Journal :
Review of Scientific Instruments
Accession number :
edsair.doi...........d0e06123a28182814fb6f2d33de5e7e6
Full Text :