Amperometric biosensor precision improvement: application to organophosphorus pesticide determination

Authors :
Margarita Stoytcheva
Gisela Montero
Velizar Gochev
Zdravka Velkova
Roumen Zlatev
Source :
Anal. Methods. 6:8232-8238
Publication Year :
2014
Publisher :
Royal Society of Chemistry (RSC), 2014.

Abstract

An approach for precision improvement in organophosphorus pesticide determination was developed, applied and characterized. The enzyme organophosphorus hydrolase was employed as a specific agent for electrode surface modification, and the pulsed amperometric detection for 66.6 ms was applied for the hydrolysis product detection. Pulsed electrochemical cleaning, i.e. oxidation at a very positive potential (+1.4 V vs. Ag, AgCl) was applied every 233 ms for 166.6 ms for elimination of the fouling effect, which degrades the determination precision. By this way a RSD of 0.41% at 40 μmol L−1p-nitrophenol determination was achieved after 10 determinations, instead of 43.93% without the pulsed cleaning application.

Details

ISSN :
17599679 and 17599660
Volume :
6
Database :
OpenAIRE
Journal :
Anal. Methods
Accession number :
edsair.doi...........cbe4cf086c4f4e778244f513dab4a27e
Full Text :
https://doi.org/10.1039/c4ay01792g