Back to Search
Start Over
Amperometric biosensor precision improvement: application to organophosphorus pesticide determination
- Source :
- Anal. Methods. 6:8232-8238
- Publication Year :
- 2014
- Publisher :
- Royal Society of Chemistry (RSC), 2014.
-
Abstract
- An approach for precision improvement in organophosphorus pesticide determination was developed, applied and characterized. The enzyme organophosphorus hydrolase was employed as a specific agent for electrode surface modification, and the pulsed amperometric detection for 66.6 ms was applied for the hydrolysis product detection. Pulsed electrochemical cleaning, i.e. oxidation at a very positive potential (+1.4 V vs. Ag, AgCl) was applied every 233 ms for 166.6 ms for elimination of the fouling effect, which degrades the determination precision. By this way a RSD of 0.41% at 40 μmol L−1p-nitrophenol determination was achieved after 10 determinations, instead of 43.93% without the pulsed cleaning application.
Details
- ISSN :
- 17599679 and 17599660
- Volume :
- 6
- Database :
- OpenAIRE
- Journal :
- Anal. Methods
- Accession number :
- edsair.doi...........cbe4cf086c4f4e778244f513dab4a27e
- Full Text :
- https://doi.org/10.1039/c4ay01792g