Back to Search Start Over

In vitro retinal imaging with full field swept source optical coherence tomography

Authors :
Boris Považay
Bernd Hofer
Wolfgang Drexler
James Fergusson
Source :
SPIE Proceedings.
Publication Year :
2010
Publisher :
SPIE, 2010.

Abstract

Weakly scattering tree shrew retina has been imaged in vitro with full field swept source optical coherence tomography, visualising multiple intraretinal layers. The system utilises a 50nm bandwidth Superlum SLD, to acheive ~8μm of axial resolution and 4μm of transversal resolution. Volumetric images of retinal tissue with dimensions of 1248x936x678μm (horizontal by vertical by axial) were recorded in two second (equivalent of 153,600 A-scans per second) with a measured signal to noise ratio of 75dB. From the 5mW of SLD optical power available, 720μW illuminates the sample, giving a power per pixel of 4.6nW, ten times less power per pixel then standard FDOCT systems. After upgrading the camera and redesigning the optical beam path, 82dB of SNR was realised.

Details

ISSN :
0277786X
Database :
OpenAIRE
Journal :
SPIE Proceedings
Accession number :
edsair.doi...........cb21db69feaebf5d75aebaa2ea4c2fb1
Full Text :
https://doi.org/10.1117/12.843254