Back to Search Start Over

Frequency stability of an optically pumped caesium-beam frequency standard at the KRISS

Authors :
Sung-Hoon Yang
Y B Kim
P S Kim
J O Kim
Cha-Hwan Oh
Ho Seong Lee
K J Baek
Source :
Metrologia. 35:25-31
Publication Year :
1998
Publisher :
IOP Publishing, 1998.

Abstract

An optically pumped caesium-beam frequency standard (known as the KRISS-1) has been constructed at the Korea Research Institute of Standards and Science (KRISS) and the first measurements of long-term frequency stability have been made. Two methods were used to make the frequency standard immune to variations in the environmental temperature: an F = 4 → F' = 3 (σ + π)-polarized laser was used as a pumping source, and the difference between two sequential error signals was used as a feedback signal in the digital servo loop. A frequency stability of σy(τ) = 2.5 × 10-14 at τ = 105 s for a temperature variation of 4 °C was achieved with the KRISS-1.

Details

ISSN :
00261394
Volume :
35
Database :
OpenAIRE
Journal :
Metrologia
Accession number :
edsair.doi...........ca399acf92d6d0f0282ccb6290a59452
Full Text :
https://doi.org/10.1088/0026-1394/35/1/5