- Aiming at the limits of continuum institutions surgery. This paper presents two modules design of continuum institutions surgical instruments which include the design of “backbone” and execution ends. We establish a kind of new snake-continuum institutions with 36 degrees of freedom which has four wire line driver. Its characteristics are better able to achieve “flexible” movement. We design a kind of the end actuator which has 4-DOF. Determine the coordinate position of its motion space. Establish a ball connection of flexible body and the kinematic model of the end actuator. Conducted a motion simulation and finite element analysis with ANSYS to the continuum institutions.Analysis shows: the maximum stress point is far less than the tensile strength and yield limit of parts. Used in minimally invasive surgery for the institution to provide a theoretical basis for the throat.