Mechanical design of a dual-cryostat instrument for a high-field pulsed magnet

Authors :
Hiroyuki Nojiri
Emil Trakhtenberg
Yasuo Narumi
Dana Capatina
Zahirul Islam
Source :
Diamond Light Source Proceedings. 1
Publication Year :
2010
Publisher :
Cambridge University Press (CUP), 2010.

Abstract

The engineering of a dual-cryostat for a pulsed-magnet instrument at the Advanced Photon Source is presented. The dual-cryostat independently cools the magnet coil (using liquid-nitrogen) and the sample (using a closed-cycle refrigerator). Liquid-nitrogen cooling may allow a repetition rate of a few minutes for peak fields near 30 T. The system is unique in that the liquid-nitrogen cryostat incorporates a double-funnel vacuum tube passing through the solenoid's bore in order to preserve the entire angular range allowed by the magnet bore for scattering studies. Second, the use of a separate refrigerator for the sample allows precise positioning of samples in the bore while minimizing magnet vibrations propagating to the sample during pulsed-field generation.