Back to Search Start Over

Separator of primary and signal electrons for very low energy SEM

Authors :
Luděk Frank
Ilona Müllerová
V. Kolařík
D. Pejchl
Source :
Czechoslovak Journal of Physics. 44:269-276
Publication Year :
1994
Publisher :
Springer Science and Business Media LLC, 1994.

Abstract

High resolution can be obtained in the scanning electron microscope (SEM) for a very low landing energy of electrons, even for that below 50 eV, when a cathode lens is used with the specimen as a cathode held at a high negative potential but the detection of signal electrons is totally different compared with classical SEM. Primary electrons with an energy of the order of tens of keV are decelerated in the field of the cathode lens to a very low landing energy and signal electrons originating in the specimen are accelerated and collimated by the same field to a narrow beam with an electron energy nearly the same as that of the primary beam. To detect these signal electrons we must deflect them from the optical axis without deteriorating the properties of the primary beam. The design of a novel type of separator of the primary and signal electrons consisting of two stages, each of them formed by the electric and magnetic crossed fields, is presented, together with calculated trajectories for both primary and signal electrons.