Back to Search Start Over

The design of respiratory monitoring system based on laser reflection for infant incubator

Authors :
Nji Raden Poespawati
Kresna Devara
Fitriyanti Nur Aisyah
Ketut Vanda Aryanthera Wisnu Nadia
Retno Wigajatri Purnamaningsih
Source :
AIP Conference Proceedings.
Publication Year :
2019
Publisher :
Author(s), 2019.

Abstract

We proposed the design of a respiration system based on laser reflection for an infant incubator that is able to continuously operate. The device consists of a laser diode (λ=650 nm; P=5 mW), a paper reflector, photodiode array as a detector, comparator, microcontroller Arduino UNO R3, and LCD. Also, there is a buzzer for anticipating emergency situations such as apnea. It utilizes the reflection of laser light on paper reflector attached to an infant simulator. Next, the reflection will be received by photodiode array and comparator which connects to the Arduino UNO R3. Then, the result was displayed on an LCD. From the experiment results, it was indicated that the system could measure respiratory rate from 10 up to 70 breaths per minute with an error of 2.357%.

Details

ISSN :
0094243X
Database :
OpenAIRE
Journal :
AIP Conference Proceedings
Accession number :
edsair.doi...........bbff0fd92a4a5018670aef9ca9513c19