Detorque Evaluation of Dental Abutment Screws after Immersion in a Fluoridated Artificial Saliva Solution
- Source :
- Journal of Prosthodontics. 22:275-281
- Publication Year :
- 2012
- Publisher :
- Wiley, 2012.
-
Abstract
- Purpose Implant-abutment connections still present failures in the oral cavity due to the loosening of mechanical integrity by detorque and corrosion of the abutment screws. The objective of this study was to evaluate the detorque of dental abutment screws before and after immersion in fluoridated solutions. Materials and Methods Five commercial implant-abutment assemblies were assessed in this investigation: (C) Conexao®, (E) Emfils®, (I) INP®, (S) SIN®, and (T) Titanium Fix®. The implants were embedded in an acrylic resin and then placed in a holding device. The abutments were first connected to the implants and torqued to 20 Ncm using a handheld torque meter. The detorque values of the abutments were evaluated after 10 minutes. After applying a second torque of 20 Ncm, implant-abutment assemblies were withdrawn every 3 hours for 12 hours in a fluoridated solution over a period of 90 days. After that period, detorque of the abutments was examined. Scanning electronic microscopy (SEM) associated to energy dispersive spectroscopy (EDS) was applied to inspect the surfaces of abutments. Results Detorque values of systems C, E, and I immersed in the fluoridated solution were significantly higher than those of the initial detorque. ANOVA demonstrated no significant differences in detorque values between designs S and T. Signs of localized corrosion could not be detected by SEM although chemical analysis by EDS showed the presence of elements involved in corrosive processes. Conclusion An increase of detorque values recorded on abutments after immersion in fluoridated artificial saliva solutions was noticed in this study. Regarding chemical analysis, such an increase of detorque can result from a corrosion layer formed between metallic surfaces at static contact in the implant-abutment joint during immersion in the fluoridated solutions.
- Subjects :
- Materials science
business.industry
Abutment
Dentistry
030206 dentistry
02 engineering and technology
021001 nanoscience & nanotechnology
Oral cavity
Corrosion
03 medical and health sciences
0302 clinical medicine
Dental Implant-Abutment Design
Dental Abutments
visual_art
visual_art.visual_art_medium
Immersion (virtual reality)
Electronic microscopy
0210 nano-technology
business
General Dentistry
Acrylic resin
Subjects
Details
- ISSN :
- 1059941X
- Volume :
- 22
- Database :
- OpenAIRE
- Journal :
- Journal of Prosthodontics
- Accession number :
- edsair.doi...........b9d9b5fd3c177162c3d6c55d637fbd01
- Full Text :
- https://doi.org/10.1111/j.1532-849x.2012.00941.x