Comparison of magnetostimulation limits for axial and transverse drive fields in MPI

Authors :
Steven M. Conolly
Emine Ulku Saritas
Elaine Y. Yu
Source :
2013 International Workshop on Magnetic Particle Imaging (IWMPI).
Publication Year :
2013
Publisher :
IEEE, 2013.

Abstract

As shown in Fig. 1b, the mean magneto stimulation thresholds for the transverse field was approximately 3 times lower than the axial field (ΔBmin = 26 mT-pp vs. 78 mT-pp) for the same set of subjects, with comparable τc values (265 μs vs. 243 μs). In addition, the percentage deviation across subjects was significantly smaller for the transverse field (13%) than the axial field (28%). For both transverse and axial fields, the data trends were well modeled by the hyperbolic amplitude vs. frequency curve, as demonstrated in Fig. 1c for a single subject.