Back to Search
Start Over
Measuring Seebeck coefficient on thin film thermoelectric materials without metallization treatment
- Source :
- 2015 10th International Microsystems, Packaging, Assembly and Circuits Technology Conference (IMPACT).
- Publication Year :
- 2015
- Publisher :
- IEEE, 2015.
-
Abstract
- This paper developed a new methodology which is a sandwich-like platform to measure Seebeck coefficient for single layer thin-film thermoelectric devices which are flexible and wearable without environmental limitation. With our new apparatus, a stably controlled temperature gradient environment is created on thin-film test samples so that the commonly used and simple probing can be employed for Seebeck coefficient measurements. This paper finally verified the accuracy and availability of the new experimental design through Seebeck coefficient measurements on a typical organic conductive material (PEDOT)
Details
- Database :
- OpenAIRE
- Journal :
- 2015 10th International Microsystems, Packaging, Assembly and Circuits Technology Conference (IMPACT)
- Accession number :
- edsair.doi...........af003b2fc9514c407e9f11a45fc9dc82
- Full Text :
- https://doi.org/10.1109/impact.2015.7365204