A Differential Pressure Approach to Spirometry
- Source :
- 2007 IEEE Biomedical Circuits and Systems Conference.
- Publication Year :
- 2007
- Publisher :
- IEEE, 2007.
-
Abstract
- This paper presents the design, implementation and measurements of a spirometer based on differential pressure sensing. A system which fulfills the last medical standard specifications has been designed exploiting the Venturi tube principle. A fully operating prototype has been tested, and data have been subsequently analyzed. Air flows up to 14 L/s can be measured with an accuracy of 0.2 L/s. Data acquisition software and user interface were developed.
Details
- Database :
- OpenAIRE
- Journal :
- 2007 IEEE Biomedical Circuits and Systems Conference
- Accession number :
- edsair.doi...........aaaec1060b52de973aee8a41b5246287
- Full Text :
- https://doi.org/10.1109/biocas.2007.4463295