Abstract

Purpose: To evaluate in vitro the maximum pressure generated in an artificial kidney model when people of different levels of strengths used various irrigation systems. Materials and Methods: Fifteen people were enrolled and divided into three groups based on their strengths. Individual strength was evaluated according to the maximum pressure each participant was able to achieve using an Encore™ Inflator. The irrigation systems evaluated were as follows: T-Flow™ Dual Port, Hiline™, continuous flow single action pumping system (SAPS™) with the system close and open, Irri-flo II™, a simple 60-mL syringe, and Peditrol™. Each irrigation system was connected to URF-V2 ureteroscope, which was inserted into an artificial kidney model. Each participant was asked to produce the maximum pressure possible with every irrigation device. Pressure was measured with the working channel (WC) empty, with a laser fiber, and a basket inside. Results: The highest pressure was achieved with the 60 mL-syringe system an...