Uitra-Flexible, High-Density Neural Electrode Probes For Reliable Multi-Region Neural Activity Monitoring
- Source :
- 2021 IEEE 34th International Conference on Micro Electro Mechanical Systems (MEMS).
- Publication Year :
- 2021
- Publisher :
- IEEE, 2021.
-
Abstract
- We designed, and fabricated an ultra-flexible nanofabricated neural probe to achieve reliable multi-region neural activities recording simultaneously in both superficial and deep structures of brain for freely moving mice. High spiking yield of up to 85.2 % has been achieved for over 8 weeks with a minimally invasive (with each shank only 0.5 mm diameter hole drill for craniotomy) in-vivo implantation. Each probe has 128 recording sites evenly distributed over 4 shanks with each shank length of 2700 microns, cross section of 72×1.5 -μm. The flexible nanofabricated neural probes were then implanted in freely moving mice, achieving both acute and chronic recordings from the neurons in the superficial region motor cortex (M1) and deep region caudate putamen (CPu). The scalable neural probe fabrication methods and high performance ultraflexible probes pave the way towards the brain-wide neural activities monitoring for freely behaving lives.
- Subjects :
- Materials science
High density
02 engineering and technology
010402 general chemistry
021001 nanoscience & nanotechnology
01 natural sciences
0104 chemical sciences
Caudate putamen
Cross section (geometry)
Neural activity
medicine.anatomical_structure
Fabrication methods
Electrode
medicine
0210 nano-technology
Biomedical engineering
Motor cortex
Subjects
Details
- Database :
- OpenAIRE
- Journal :
- 2021 IEEE 34th International Conference on Micro Electro Mechanical Systems (MEMS)
- Accession number :
- edsair.doi...........a4c88bdd0ad9c242f58872b98177e0eb