Back to Search Start Over

LaBr3(Ce) gamma-ray detector for neutron capture therapy

Authors :
Alexander E. Shustov
Konstantin F. Vlasik
A.S. Novikov
D. V. Petrenko
M Smirnova
Valery V. Dmitrenko
S. Z. Shmurak
A. D. Orlov
E Shmanin
A V Galavanov
Sergey E. Ulin
Z. M. Uteshev
Source :
Journal of Physics: Conference Series. 675:042050
Publication Year :
2016
Publisher :
IOP Publishing, 2016.

Abstract

Results of developing of a gamma-ray detector based on LaBr3(Ce) scintillation crystal for neutron capture therapy are presented. An energy resolution of the detector measured by photomultiplier tube Hamamatsu R6233-100 is showed. It was 2.93% for gamma line 662 keV from a source 137Cs. For radiative capture gamma line of isotope 10B (478 keV) and annihilation line (511 keV) the values were 3.33 and 3.24% respectively. Data analysis of gamma spectra for an estimation of energy resolution threshold required for visual identification gamma lines 478 and 511 keV was made.

Details

ISSN :
17426596 and 17426588
Volume :
675
Database :
OpenAIRE
Journal :
Journal of Physics: Conference Series
Accession number :
edsair.doi...........9ec3219f30e5e7a167f652b8663b2540