Analytical Methods Validation of Retinoic Acid and Hydroquinone Using Ultra High Performance Liquid Chromatography in Medicinal Cream

Authors :
Linda Mazroatul
Annisa Ananda Ulya
Supandi Supandi
Umar Mansur
Source :
Pharmaceutical and Biomedical Sciences Journal (PBSJ). 1
Publication Year :
2019
Publisher :
LP2M Universitas Islam Negeri (UIN) Syarif Hidayatullah Jakarta, 2019.

Abstract

Hydroquinone and retinoic acid have been widely used as whitening agents in cosmetic creams and have a negative impact if used without a medical prescription. In this study, development of analytical method of retinoic acid (REA) and hydroquinone (HYQ) were carried out using Ultra High Performance Liquid Chromatography (UHPLC) with a PDA detector, Phenomenex Luna® ODS / C18 column (150 mmx4.6 mm), methanol-acetic acid 0.1% (85:15) as mobile phase, flow rate of 1.0 mL/min and detected at 295 nm for HYQ and 341 nm for REA. The result of the validation method showed a retention time of HYQ at 1.55 minutes and REA at 3.73 minutes. Validation parameters performed are selectivity, range, linearity, LOD, LOQ, accuracy, and precision. Linear detector response in the concentration range of 75-250 ppm with a calibration curve equation. The calibration curve equation for HYQ is y = 0.4607x + 1.9014 (r = 0.999), mean while REA is y = 2.3138x + 21.0429 (r = 0.999). LOD and LOQ values of HYQ are 1.615 ppm; 13.98 ppm respectively and REA 1.02 ppm and 3.11 ppm, respectively. The precision (% RSD) ≤ ± 2% and accuracy (% recovery) in the range of 99.0% - 100.6%. This method is valid to be used to analyze HYQ and REA in commercial cosmetic cream samples.

Details

ISSN :
27156702 and 27152979
Volume :
1
Database :
OpenAIRE
Journal :
Pharmaceutical and Biomedical Sciences Journal (PBSJ)
Accession number :
edsair.doi...........9e0d3f95d0528f3952450f1d0fc48a67
Full Text :
https://doi.org/10.15408/pbsj.v1i1.12689