Back to Search Start Over

Identification of environmental staphylococci and coliforms by the Vitek Auto Microbic System

Authors :
M. Durrer
T. Kleiss
Y. Fuhrer
P. Deberghes
J.-L. Cordier
Source :
Food Microbiology. 12:199-202
Publication Year :
1995
Publisher :
Elsevier BV, 1995.

Abstract

The Vitek Auto Microbic System (AMS) automated identification device was used in rapid identification of staphylococci and coliforms originating from food factory environments. From a total of 3324 isolates, an average of 76% were identified with a confidence > 50% (staphylococci: 61%, coliforms: 82%). AMS identifications of presumed staphylococci did not match well (24%) with ATB 32 STAPH identifications, but coliforms matched for 80% of the isolates compared with the API 20 E system. The differences observed between the AMS and ATB 32 STAPH in identifications could be due to inadequate isolation media, insufficient isolate purification, incorrect pre-screening of isolates or an incompetent AMS database. For coliforms, the overall performance proved sufficient for rapid screening of isolates from food factory bacterial ecology studies.

Details

ISSN :
07400020
Volume :
12
Database :
OpenAIRE
Journal :
Food Microbiology
Accession number :
edsair.doi...........9b5fd57cca78813df456017ffb0f57f1
Full Text :
https://doi.org/10.1016/s0740-0020(95)80098-0