Back to Search
Start Over
Identification of environmental staphylococci and coliforms by the Vitek Auto Microbic System
- Source :
- Food Microbiology. 12:199-202
- Publication Year :
- 1995
- Publisher :
- Elsevier BV, 1995.
-
Abstract
- The Vitek Auto Microbic System (AMS) automated identification device was used in rapid identification of staphylococci and coliforms originating from food factory environments. From a total of 3324 isolates, an average of 76% were identified with a confidence > 50% (staphylococci: 61%, coliforms: 82%). AMS identifications of presumed staphylococci did not match well (24%) with ATB 32 STAPH identifications, but coliforms matched for 80% of the isolates compared with the API 20 E system. The differences observed between the AMS and ATB 32 STAPH in identifications could be due to inadequate isolation media, insufficient isolate purification, incorrect pre-screening of isolates or an incompetent AMS database. For coliforms, the overall performance proved sufficient for rapid screening of isolates from food factory bacterial ecology studies.
Details
- ISSN :
- 07400020
- Volume :
- 12
- Database :
- OpenAIRE
- Journal :
- Food Microbiology
- Accession number :
- edsair.doi...........9b5fd57cca78813df456017ffb0f57f1
- Full Text :
- https://doi.org/10.1016/s0740-0020(95)80098-0