Back to Search Start Over

Energy calibration of gamma spectra in plastic scintillators using Compton kinematics

Authors :
Edward R. Siciliano
Denis M. Strachan
James H. Ely
Satoru T. Yokuda
Richard T. Kouzes
John E. Schweppe
Source :
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 594:232-243
Publication Year :
2008
Publisher :
Elsevier BV, 2008.

Abstract

This paper describes a simple and practicable method for assigning energy values to gamma-ray pulse-height distributions measured with polyvinyl toluene (PVT) based detectors. It is based upon the characteristic shape of simulated spectra in the region of maximum energy deposition resulting from a single Compton scattering. The validity of this method is demonstrated by applying it to a set of measured NaI(Tl) spectra, and comparing those results to the standard photopeak method of calibrating the same spectra. The method is then applied to a set of measured PVT spectra. It can now be applied to calibration of fielded PVT detectors using two selected gamma-ray sources, which was the initial motivation for this study.

Details

ISSN :
01689002
Volume :
594
Database :
OpenAIRE
Journal :
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Accession number :
edsair.doi...........9adf39476454e8ffe86b38b4e4b0d54a
Full Text :
https://doi.org/10.1016/j.nima.2008.06.031