Back to Search Start Over

Range sensitive telescope detector

Authors :
S. Micek
K. Grotowski
Z. Sosin
T. Kozik
Source :
Nuclear Instruments and Methods in Physics Research. 224:454-460
Publication Year :
1984
Publisher :
Elsevier BV, 1984.

Abstract

A telescope for the identification of heavy ions over an extended energy range is described. The energy loss Δ E is measured with an ionisation chamber supplied with the range-sensitive charge collection system. The particle residual energy E can be measured with a solid state detector. Two modes of operation are available: a standard ΔEt − E identification, or a charge integration along the track with a continuous resistivity anode. By a proper formation of this resistivity, one can obtain an identification signal proportional to the Z value in a large range of incident energies.

Details

ISSN :
01675087
Volume :
224
Database :
OpenAIRE
Journal :
Nuclear Instruments and Methods in Physics Research
Accession number :
edsair.doi...........90f88c449a7edc9001e7d0645c5c4e4f
Full Text :
https://doi.org/10.1016/0167-5087(84)90037-1