Back to Search Start Over

High-resolution pluronic-filled microchip CE-SSCP analysis system via channel width control

Authors :
Giyoung Shin
Dong-Kyun Kim
Gyoo Yeol Jung
Junsang Doh
Nam Ki Lee
Daeyeon Lee
Source :
ELECTROPHORESIS. 37:676-679
Publication Year :
2015
Publisher :
Wiley, 2015.

Abstract

Although the resolution of CE-SSCP has been significantly improved by using a poly(ethyleneoxide)-poly(propyleneoxide)-poly(ethyleneoxide) (PEO-PPO-PEO; Pluronic(®)) triblock copolymer as a separation medium, CE-SSCP on a microchip format is not widely applicable because their resolution is limited by short channel length. Therefore, a strategy to improve the resolution in channels of limited lengths is highly required for enabling microchip-based CE-SSCP. In this study, we developed a high-resolution CE-SSCP microchip system by controlling the width of the pluronic-filled channel. We tested four different channel widths of 180, 240, 300, and 400 μm, and found that 300 μm showed the highest resolution in the separation of two pathogen specific markers. Potential applications of our method in various genetic analyses were also shown by using SNP markers for spinal muscular atrophy.

Details

ISSN :
01730835
Volume :
37
Database :
OpenAIRE
Journal :
ELECTROPHORESIS
Accession number :
edsair.doi...........8cee52f43fa6d70e2204c2acc51c3720
Full Text :
https://doi.org/10.1002/elps.201500427