Back to Search Start Over

Analysis of IAEA environmental samples for plutonium and uranium by ICP/MS in support of international safeguards

Authors :
Khris B. Olsen
Stephanie J. Garofoli
May-Lin P. Thomas
Orville T. Farmer
Source :
Journal of Radioanalytical and Nuclear Chemistry. 276:489-492
Publication Year :
2008
Publisher :
Springer Science and Business Media LLC, 2008.

Abstract

A method for the separation and determination of total and isotopic uranium and plutonium by ICP/MS was developed for IAEA samples on cellulose-based media. Preparation of the IAEA samples involved a series of redox chemistries and separations using TRU® resin (Eichrom). The sample introduction system, an APEX nebulizer (Elemental Scientific, Inc.), provided enhanced nebulization for a several-fold increase in sensitivity and reduction in background. Application of mass bias (ALPHA) correction factors greatly improved the precision of the data. By combining the enhancements of chemical separation, instrumentation and data processing, detection levels for uranium and plutonium approached high attogram levels.