Comparison of Continuous, Stat, and Intermittent Bolus Cardiac Output Measurements during Cardiac Surgery
- Source :
- Korean Journal of Anesthesiology. 54:603
- Publication Year :
- 2008
- Publisher :
- The Korean Society of Anesthesiologists, 2008.
-
Abstract
- (46.2%) for CCI and 0.09 ± 0.60 L/min/m 2 (44.0%) for SCI, compared with ICI measurements. Conclusions: We found that the agreement of CCI, SCI, and ICI was poor. This disagreement increased during periods of open thorax and open pericardium when compared to the period of closed thorax. Continuous CO measurements through pulmonary artery catheter should be interpreted with caution during OPCAB. (Korean J Anesthesiol 2008; 54: 603∼8)
- Subjects :
- Thorax
medicine.medical_specialty
Cardiac output
animal structures
business.industry
medicine.medical_treatment
Pulmonary artery catheter
Cardiac surgery
Anesthesiology and Pain Medicine
medicine.anatomical_structure
Internal medicine
Anesthesia
Intermittent bolus
medicine
Cardiology
Pericardium
business