Capillary supercritical fluid chromatography with dual-flame photometric detection

Authors :
Randy. Bolick
Ed D. Lee
Karin E. Markides
Milton L. Lee
Source :
Analytical Chemistry. 58:740-743
Publication Year :
1986
Publisher :
American Chemical Society (ACS), 1986.

Abstract

A dual-flame photometric detector was modified to be used in a capillary supercritical fluid chromatographic system. The detection limit was found to be 25 ng for the sulfur mode and 0.5 ng for the phosphorus mode when supercritical carbon dioxide was used as the mobile phase. The advantages and limitations are discussed and demonstrated with several applications. 18 references, 6 figures.