X-ray microbeam with sputtered-sliced Fresnel zone plate at SPring-8 undulator beamline
- Source :
- Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. :951-953
- Publication Year :
- 2001
- Publisher :
- Elsevier BV, 2001.
-
Abstract
- A hard X-ray microbeam created from a sputtered-sliced Fresnel zone plate has been tested at SPring-8 undulator beamline. Focusing properties are evaluated in the X-ray wavelength regions of 0.15– 1.5 A . The measured focal beam size is about 0.6 μm at an X-ray wavelength of 1.4 A . In a scanning microscopy experiment, resolution-test-patterns with 0.2 μm structure are resolved at an X-ray wavelength of 0.45 A .