Efficacy and Usability of a Novel Barrier Device for Preventing Injection Port Contamination

Authors :
Ronak Shah
Anjan M. Shah
Daniel S. Katz
David B. Wax
Hung-Mo Lin
Source :
Anesthesia & Analgesia. 130:e45-e48
Publication Year :
2020
Publisher :
Ovid Technologies (Wolters Kluwer Health), 2020.

Abstract

Contamination of intravenous (IV) ports and stopcocks has been associated with postoperative infections. We tested the usability and efficacy of a novel passive shielding device to prevent such contamination even in the absence of hand hygiene or port disinfection. In a desktop setting with deliberately contaminated hands, qualitative port contamination was detected after 5/60 (8.3%; 95% confidence interval [CI], 2.8-18.4) control port injections versus 0/60 (0%; 95% CI, 0-6.0) shielded injections (P = .025). In clinical simulations with a quantitative bioburden assay (measured in relative light units [RLUs]), median (interquartile range [IQR]) postsimulation bioburden was 46 (32-53) vs 27 (21-42) RLU for the control versus intervention groups (P = .036), yielding a median shift of -13 RLU (95% CI, -2 to -26) in favor of the shielding. Usability of the device was acceptable to practitioners.