Geometrical properties of a digital beam attenuator system

Authors :
Walter W. Peppler
Charles A. Mistretta
Bruce H. Hasegawa
James T. Dobbins
Shaikh Naimuddin
Source :
Medical Physics. 14:314-321
Publication Year :
1987
Publisher :
Wiley, 1987.

Abstract

A digital beam attenuator system has been developed to automatically generate patient‐specific compensating filters for chest radiography. An initial low‐dose test image is used to generate the attenuator, which is fabricated by overprinting multiple layers of a heavy‐metal material onto a nonattenuating substrate. The attenuator is subsequently inserted into the x‐ray beam for a final compensated radiograph. The effects of focal spot blurring and limited attenuator resolution result in the final compensated image containing only high‐spatial frequency information. The frequency response of the process is not strictly describable by a modulation transfer function, but an approximation of the frequencies remaining in the compensated image is obtained for low‐contrast conditions. It is found that a 4×4 blurring function on the original 64×64 test image is required for the attenuator to give appropriate compensated image appearance. A proposed attenuator printing scheme prints the attenuator in a 16×16 matrix, staggering successively printed layers to achieve the required 64×64 sampling with appropriate blurring. The resulting compensated image has good anatomical definition and contains a frequency response similar to that obtained by compensation techniques being investigated by Plewes and Sorenson.

Details

ISSN :
00942405
Volume :
14
Database :
OpenAIRE
Journal :
Medical Physics
Accession number :
edsair.doi...........839385a73a2e47bcc8a491e0c6e569a3
Full Text :
https://doi.org/10.1118/1.596086