Back to Search
Start Over
Dose per Pulse Monitoring of MeV Photon Beams
- Source :
- ATHENA Research Book, Volume 1 ISBN: 9789612866587
- Publication Year :
- 2022
- Publisher :
- University of Maribor Press, 2022.
-
Abstract
- Modern dynamic radiotherapies in cancer treatment are oriented to release high dose values in the shortest irradiation time. For these techniques, fast detectors and dedicated conditioning electronics will play a fundamental role for beam diagnostics to verify the necessary quality assurance requirements of treatments. This work describes a measurement system based on a single-crystal synthetic diamond able to monitor the dose released by each X-ray pulse generated by a medical LINAC. A gated-integration method, performed by a specifically designed electronics, has been used to measure the charge photogenerated in diamond by each impinging X-pulse. Exploiting the synchronization signal available in a LINAC apparatus, our system allows performing pulse-beam diagnostics with an acquisition time limited to the period around each pulse. Therefore, the proposed solution represents an effective tool for real-time dosimetry in modern radiotherapy applications and assures a better accuracy in comparison to conventional electrometer-based measurement systems that require longer acquisition periods.
Details
- ISBN :
- 978-961-286-658-7
- ISBNs :
- 9789612866587
- Database :
- OpenAIRE
- Journal :
- ATHENA Research Book, Volume 1 ISBN: 9789612866587
- Accession number :
- edsair.doi...........81927d3d0ea39193a804ad56b9c86a71
- Full Text :
- https://doi.org/10.18690/um.3.2022.16