Back to Search Start Over

Flow-injection pulse amperometric detection based on ion transfer across a water-plasticized polymeric membrane interface for the determination of imipramine

Authors :
Antonio José Meseguer Gil
J. Ortuño
Concepción Sánchez-Pedreño
Source :
Sensors and Actuators B: Chemical. 122:369-374
Publication Year :
2007
Publisher :
Elsevier BV, 2007.

Abstract

A flow-injection pulse amperometric method for determining imipramine, based on ion transfer across a plasticized poly(vinyl chloride) (PVC) membrane, was developed. A four-electrode potentiostat with ohmic drop compensation was used, while a flow-through cell incorporated the four electrodes and the membrane, which contained tetrabutylammonium tetraphenylborate. The influence of the applied potential, the flow-injection variables and of the pH on the determination of imipramine was studied. In the selected conditions, a linear relationship between current peak height and imipramine concentration over a range of 1 × 10 −6 to 8 × 10 −5 M imipramine was found. Good repeatability and between-day reproducibility were obtained. No interference was observed on the part of some common ions and pharmaceutical excipients. The method proposed was applied satisfactorily to the determination of imipramine in pharmaceuticals and human serum.

Details

ISSN :
09254005
Volume :
122
Database :
OpenAIRE
Journal :
Sensors and Actuators B: Chemical
Accession number :