Back to Search Start Over

Combining Monte Carlo simulations and dosimetry measurements for process control in the Tunisian Cobalt-60 irradiator after three half lives of the source

Authors :
Shayma Mohamed Abdallah
Leila Ounalli
N. Reguigui
Maroua Bhar
Ammar Bouabidi
Arbi Mejri
K. Manai
Source :
Nuclear Science and Techniques. 28
Publication Year :
2017
Publisher :
Springer Science and Business Media LLC, 2017.

Abstract

To check the dose uniformity and to determine the efficiency of medical devices sterilization by gamma irradiation after three half lives of the source, calculations of the absorbed dose were carried out. Monte Carlo simulations and dosimetry measurements, were established to study the radiation processing quality control. An isodose chart was created by GEANT4 Monte Carlo code to evaluate the absorbed dose rate uniformity inside the irradiation room from the year of the installation until the year of the source reload. The dose uniformity ratio (DUR) is deduced from maximum and minimum experimental doses in medical devices after three half lives of the source.

Details

ISSN :
22103147 and 10018042
Volume :
28
Database :
OpenAIRE
Journal :
Nuclear Science and Techniques
Accession number :
edsair.doi...........7b706d503b687bebfefcde68795debea
Full Text :
https://doi.org/10.1007/s41365-017-0289-5