Back to Search
Start Over
Combining Monte Carlo simulations and dosimetry measurements for process control in the Tunisian Cobalt-60 irradiator after three half lives of the source
- Source :
- Nuclear Science and Techniques. 28
- Publication Year :
- 2017
- Publisher :
- Springer Science and Business Media LLC, 2017.
-
Abstract
- To check the dose uniformity and to determine the efficiency of medical devices sterilization by gamma irradiation after three half lives of the source, calculations of the absorbed dose were carried out. Monte Carlo simulations and dosimetry measurements, were established to study the radiation processing quality control. An isodose chart was created by GEANT4 Monte Carlo code to evaluate the absorbed dose rate uniformity inside the irradiation room from the year of the installation until the year of the source reload. The dose uniformity ratio (DUR) is deduced from maximum and minimum experimental doses in medical devices after three half lives of the source.
- Subjects :
- Nuclear and High Energy Physics
medicine.medical_specialty
Materials science
010308 nuclear & particles physics
Nuclear engineering
Monte Carlo method
01 natural sciences
030218 nuclear medicine & medical imaging
Dose uniformity
03 medical and health sciences
0302 clinical medicine
Nuclear Energy and Engineering
Absorbed dose
0103 physical sciences
medicine
Process control
Dosimetry
Medical physics
Irradiation
Absorbed dose rate
Cobalt-60
Subjects
Details
- ISSN :
- 22103147 and 10018042
- Volume :
- 28
- Database :
- OpenAIRE
- Journal :
- Nuclear Science and Techniques
- Accession number :
- edsair.doi...........7b706d503b687bebfefcde68795debea
- Full Text :
- https://doi.org/10.1007/s41365-017-0289-5