Back to Search
Start Over
Characterization of a deep depletion, back-illuminated charge-coupled device in the x-ray range
- Source :
- Review of Scientific Instruments. 76:116101
- Publication Year :
- 2005
- Publisher :
- AIP Publishing, 2005.
-
Abstract
- In this Note we report on the characterization of a “deep depletion” charge-coupled device (CCD) in the keV range. Measurement of quantum efficiency for different energies from 4.5 until 27keV is given. A measurement of the capability of the CCD to detect single photons is also reported together with a determination of the spatial and spectral resolution. A Monte Carlo simulation was used to give a quantitative comparison with the measured spatial resolution.
Details
- ISSN :
- 10897623 and 00346748
- Volume :
- 76
- Database :
- OpenAIRE
- Journal :
- Review of Scientific Instruments
- Accession number :
- edsair.doi...........7a826ac4c6ffd018b879c936b8c3457b
- Full Text :
- https://doi.org/10.1063/1.2093767