SU-E-T-507: Interfractional Variation of Fiducial Marker Position During HDR Brachytherapy with Cervical Interstitial Needle Template

Authors :
Ivan A. Brezovich
Rex A. Cardan
Richard A. Popple
Sui Shen
Xizeng Wu
J Araujo
L Burnett
Jun Duan
Robert Y. Kim
S Benhabib
Source :
Medical Physics. 41:343-343
Publication Year :
2014
Publisher :
Wiley, 2014.

Abstract

Purpose: HDR brachytherapy using interstitial needle template for cervical cancer is commonly delivered in 4-5 fractions. Routine verification of needle positions before each fraction is often based on radiographic imaging of implanted fiducial markers. The current study evaluated interfractional displacement of implanted fiducial markers using CT images. Methods: 9 sequential patients with cervical interstitial needle implants were evaluated. The superior and inferior borders of the target volumes were defined by fiducial markers in planning CT. The implant position was verified with kV orthogonal images before each fraction. A second CT was acquired prior 3rd fraction (one or 2 days post planning CT). Distances from inferior and superior fiducial markers to pubic symphysis plane (perpendicular to vaginal obtulator)were measured. Distance from needle tip of a reference needle (next to the inferior marker) to the pubic symphysis plane was also determined. The difference in fiducial marker distance or needle tip distance between planning CT and CT prior 3rd fraction were measured to assess markers migration and needle displacement. Results: The mean inferior marker displacement was 4.5 mm and ranged 0.9 to 11.3 mm. The mean superior marker displacement was 2.7 mm and ranged 0 to 10.4 mm. There was a good association between inferior and superior marker displacement (r=0.95). Mean averaged inferior and superior marker displacement was 3.3 mm and ranged from 0.1 to 10.9 mm, with a standard deviation of 3.2 mm. The mean needle displacement was 5.6 mm and ranged 0.2 to 15.6 mm. Needle displacements were reduced (p

Details

ISSN :
00942405
Volume :
41
Database :
OpenAIRE
Journal :
Medical Physics
Accession number :
edsair.doi...........760173edc672365fb844f672ce0f339c
Full Text :
https://doi.org/10.1118/1.4888840