Performance evaluation of a new Stago ® automated haemostasis analyser: The STA R Max ® 2
- Source :
- International Journal of Laboratory Hematology. 41:731-737
- Publication Year :
- 2019
- Publisher :
- Wiley, 2019.
-
Abstract
- Introduction The STA R Max® 2 is a new coagulation analyser developed by Diagnostica Stago, able to perform clotting, chromogenic and immuno-turbidimetric tests. A pre-analytical module build into the cap-piercing needle performs the sample integrity verification (sample tube filling and measurement of haemolysis, icterus, lipaemia). The STA R Max® 2 analyser incorporates an accreditation program tools to assist technical validation of the analyser. We assessed the analytical performance of the STA R Max® 2. Materials and methods The following tests were assessed: prothrombin time, activated partial thromboplastin time, thrombin time, fibrinogen, factor V (FV), antithrombin (AT), D-dimers (DDI) and von Willebrand factor antigen. The assay precisions were assessed using fresh plasma samples or internal quality controls. An inter-analyser comparison was performed with a STA-R Evolution® analyser or, for the FV, with a BCS® XP System. Haemolysis and icterus detection were also verified. Results For the intra-assay precision, the coefficients of variation (CV%) were all less than 5% and for DDI, the standard deviation (SD) was less than 0.1. For the inter-assay study, all CV% were less than 5%, with the exception of FV and AT (FV: 6.68% and 5.27%; AT: 7% and 12.14% for normal and pathological values, respectively). SD was less than 0.1 for DDI. The inter-analyser comparison demonstrated good results. Haemolysis and icterus were detected correctly for all our assessed samples. Conclusion According to our methods validation's recommendations, the results demonstrated a good technical and analytical performance of the STA R Max® 2 analysers for the tests assessed.
- Subjects :
- Prothrombin time
medicine.diagnostic_test
business.industry
Biochemistry (medical)
Clinical Biochemistry
Antithrombin
Analyser
Hematology
General Medicine
030204 cardiovascular system & hematology
Thrombin time
Haemolysis
Internal quality
Von Willebrand factor Antigen
03 medical and health sciences
0302 clinical medicine
medicine
Nuclear medicine
business
030215 immunology
medicine.drug
Partial thromboplastin time
Subjects
Details
- ISSN :
- 1751553X and 17515521
- Volume :
- 41
- Database :
- OpenAIRE
- Journal :
- International Journal of Laboratory Hematology
- Accession number :
- edsair.doi...........75da26cb424ae4792b398eedbaafe923
- Full Text :
- https://doi.org/10.1111/ijlh.13100