Back to Search Start Over

A STUDY OF THE FLOW CHARACTERISTIC AROUND THE SELF-CLEANING HEART VALVE UNDER PULSATILE CONDITIONS

Authors :
Y. Haggag
E. Bilgen
Publication Year :
1981
Publisher :
Elsevier, 1981.

Abstract

The self-cleaning prosthetic heart valve was tested under pulsatile flow conditions. The velocity field in the vicinity of the valve was investigated using a laser doppler anemometer. The areas of stagnation are minimum and the wall shear rate is in a physiologically acceptable range for a prosthetic device.

Details

Database :
OpenAIRE
Accession number :
edsair.doi...........7504415edb50bc7e1ae6ccfc2ac53167
Full Text :
https://doi.org/10.1016/b978-0-08-027207-8.50019-9