Accurate Measurement on Cross Sectional Area of Single Filaments Using a Desktop SEM

Authors :
Xi Ying Yang
Youqing Fei
Ou Yang Ting
Source :
Advanced Materials Research. 1083:111-117
Publication Year :
2015
Publisher :
Trans Tech Publications, Ltd., 2015.

Abstract

Cross sectional area of single filaments, a key parameter to characterize fiber properties, was experimentally studied using a desktop scanning electron microscope. Three different methods are employed based on the pixel area, averaged diameter and single diameter measurements, respectively. Results have shown that all three methods can achieve accurate measurement results once the axis of fiber sample is kept parallel to the electron beam. Significant errors are generated for the fiber samples with their axis tilted, which may frequently occur as a sample prepared. For circular fibers, a single diameter measured from tilted fibers is sufficient to determine their cross sectional area at high precision with COV values below 1.6%. By selecting an appropriate method, a desktop SEM can serve as a convenient and powerful tool for accurate determination of cross sectional area as well as morphological observation.

Details

ISSN :
16628985
Volume :
1083
Database :
OpenAIRE
Journal :
Advanced Materials Research
Accession number :
edsair.doi...........74b183d86067098c2cdca4ec65af865f
Full Text :
https://doi.org/10.4028/www.scientific.net/amr.1083.111