Back to Search Start Over

SU-E-J-244: Validation of a 6D-Robotic-Couch and Image Guidance Radiation Therapy (IGRT) System for Localization in World's First Single-Room Proton Therapy System

Authors :
Eric E. Klein
K Baker
R Cessac
K Grantham
Sreekrishna Goddu
L Rankine
Lakshmi Santanam
Source :
Medical Physics. 41:214-214
Publication Year :
2014
Publisher :
Wiley, 2014.

Abstract

Purpose: The Mevion S250 proton therapy unit is equipped with a 6D-robotic couch and IGRT system (Verity). The patient alignment process allows corrections in six degrees of freedom: translation {x,y,z}, pitch, roll, and yaw {θ,ϑ,ψ}. Geometric accuracy of couch corrections and imaging vs. radiation isocenter coincidence were quantified before clinical implementation. Methods: A commercial phantom with sixteen 2mm tungsten BBs was rigidly couch-mounted and imaged with CT. Seventeen rigid translations/rotations of known magnitude were digitally applied to the original CT image using commercial software, validated with Varian OBI system. For each altered image, phantom was mounted on robotic couch in original position, then Verity 2D:2D match (PA-LLAT) was performed using DRRs from altered images. Corrections were recorded and applied, phantom was imaged a second time and residual corrections recorded. Physical measurements verified that applied couch corrections coincided with both physical couch shifts/rotations and known CT image translations/rotations. Additionally, image vs. radiation isocenter coicidence was quantified over couch treatment angles (±90° from setup position) using radiochromic film and an image-guided couch star-shot. Posterior-anterior and left-lateral kV radiographs were taken before each beam was delivered to verify imaging/radiation isocentricity. Results: Verity suggested couch corrections and known CT shifts/rotations agreed within ±1mm (average: Δ lat=0.5mm; Δ vert=0.4mm; Δ long=0.3mm) and ± 0.4° (average: Δ pitch=0.24° Δ roll=0.01°; Δ yaw=0.10°). Physical couch measurements and Verity applied corrections agreed within ± 1mm (average: Δlat=0.5mm; Δvert=0.4mm; Δlong=0.2mm) and ±0.2° (average: Δpitch=0.03°; Δ roll=0.04°; Δ yaw=0.04°). The directionality of all translations and rotations were qualitatively verified. The image vs. radiation isocenter coincidence was

Details

ISSN :
00942405
Volume :
41
Database :
OpenAIRE
Journal :
Medical Physics
Accession number :
edsair.doi...........720e79b1570f563b4f8d6d7c36c6ffc1
Full Text :
https://doi.org/10.1118/1.4888297