Back to Search Start Over

Quantitative Method of Measuring Spot Size of Microfocus X-ray Generator

Authors :
Kazuto Hasuike
Hiroyuki Uchida
Hiroshi Tsunemi
Ken'ichi Torii
Source :
Japanese Journal of Applied Physics. 45:5277-5279
Publication Year :
2006
Publisher :
IOP Publishing, 2006.

Abstract

Microfocus X-ray generators have been extensively used for magnification radiography. In these applications, the sharpness of an image depends on the X-ray source size. Therefore, it is important to evaluate and optimize the X-ray spot quantitatively. Although several methods have been devised to date, some require complicated systems whereas others are not fully quantitative. Here, we introduce a simple, practical, and quantitative method that makes use of the penumbra of a knife edge. Because the method is very simple, it can be readily integrated as an auto focusing system of a focal spot.

Details

ISSN :
13474065 and 00214922
Volume :
45
Database :
OpenAIRE
Journal :
Japanese Journal of Applied Physics
Accession number :
edsair.doi...........6fdfe94717ef08c1c165dade82df135a
Full Text :
https://doi.org/10.1143/jjap.45.5277