Implementation on the Portable Blood Gas Analyzer and Performance Estimation

Authors :
Do-Un Jeong
Gye-Rok Jeon
Gil-Jung Kim
Jin-Woo Bae
Yoon-Bo Sim
Source :
Journal of Sensor Science and Technology. 12:34-43
Publication Year :
2003

Abstract

In this study, we implemented the measurement of pH, , of the arterial blood on a portable blood gas analysis system. This system is consist of two parts of hardware and software. The hardware part is divided into a fluidic mechanism and an electronic circuit unit. The system program is composed of operating, washing, correcting, and measuring routines. Both of 1-point and 2-point calibration schemes were used to enhance the accuracy of the measurement. In order to evaluate the performance of the developed system, we measured and performed statistical analysis on the characteristics of the sensing electrode response. As a result, coefficient variation was within 1.12, and maximum error was within 1.298%. We confirmed development possibility of portable blood gas analyzer.