Rapid high-resolution T1 mapping using a highly accelerated radial steady-state free-precession technique
- Source :
- Journal of Magnetic Resonance Imaging. 49:239-252
- Publication Year :
- 2018
- Publisher :
- Wiley, 2018.
-
Abstract
- Background T1 mapping is often used in some clinical protocols. Existing techniques are limited in slice coverage, and/or spatial-temporal resolution, or require long acquisitions. Here we present a multi-slice inversion-recovery (IR) radial steady-state free precession (radSSFP) pulse sequence combined with a principal component (PC) based reconstruction that overcomes these limitations. Purpose To develop a fast technique for multi-slice high-resolution T1 mapping. Study type Technical efficacy study done prospectively. Phantom/subjects IR-radSSFP was tested in phantoms, five healthy volunteers, and four patients with abdominal lesions. Field strength/sequence IR-radSSFP was implemented at 3T. Assessment Computer simulations were performed to optimize the flip angle for T1 estimation; testing was done in phantoms using as reference an IR spin-echo pulse sequence. T1 mapping with IR-radSSFP was also assessed in vivo (brain and abdomen) and T1 values were compared with literature. T1 maps were also compared with a radial IR-FLASH technique. Statistical tests A two-tailed t-test was used to compare T1 values in phantoms. A repeatability study was carried out in vivo using Bland-Altman analysis. Results Simulations and phantom experiments showed that a flip angle of 20˚ was optimal for T1 mapping. When comparing single to multi-slice experiments in phantoms there were no significant differences between the means T1 values (P = 0.0475). In vivo results show that T1 maps with spatial resolution as high as 0.69 mm × 0.69 mm × 2.00 mm (brain) and 0.83 mm × 0.83 mm × 3.00 mm (abdomen) can be generated for 84 brain slices in 3 min and 10 abdominal slices in a breath-hold; T1 values were comparable to those reported in literature. The coefficients of variation from the repeatability study were 1.7% for brain and 2.5-2.7% in the abdomen. Data conclusion A multi-slice IR-radSSFP technique combined with a PC-based reconstruction was demonstrated for higher resolution T1 mapping. This technique is fast, motion-insensitive and yields repeatable T1 values comparable to those in literature. Level of evidence 2 Technical Efficacy: Stage 1 J. Magn. Reson. Imaging 2019;49:239-252.
- Subjects :
- Materials science
Resolution (electron density)
Field strength
Pulse sequence
Repeatability
Imaging phantom
030218 nuclear medicine & medical imaging
03 medical and health sciences
0302 clinical medicine
Compressed sensing
Flip angle
Radiology, Nuclear Medicine and imaging
Image resolution
Biomedical engineering
Subjects
Details
- ISSN :
- 10531807
- Volume :
- 49
- Database :
- OpenAIRE
- Journal :
- Journal of Magnetic Resonance Imaging
- Accession number :
- edsair.doi...........69def746c8a67e889aa467840b8526aa