Impedimetric cardiac biomarker determination in serum mediated by epoxy and hydroxyl of reduced graphene oxide on gold array microelectrodes

Authors :
Steven Taniselass
M. K. Md Arshad
Mohamad Faris Mohamad Fathil
C. Ibau
Subash C. B. Gopinath
Periasamy Anbu
Source :
Microchimica Acta. 188
Publication Year :
2021
Publisher :
Springer Science and Business Media LLC, 2021.

Abstract

A label-free chemical bonding strategy mediated by reduced graphene oxide (rGO) basal plane functional groups has been developed for cardiac Troponin I (cTnI) detection. Four different chemical strategies on respective electrode sensing surface were precedingly examined using electrochemical impedance spectroscopy. The impedimetric assessment was carried out by sweeping frequency at the range 0.1–500 kHz perturbated at a small amplitude of AC voltage (25 mV). The chemical strategy-4 denoted as S-4 shows a significant analytical performance on cTnI detection in spiked buffer and human serum, whereby the pre-mixture of rGO and (3-Aminopropyl)triethoxysilane (APTES) creates a large number of amine sites (−NH2), which significantly enhanced the antibody immobilization without excessive functionalization. The as-fabricated immunosensor exhibited an ultra-low limit of detection of 6.3 ag mL−1 and the lowest antigen concentration measured was at 10 ag mL−1. The immunosensor showed a linear and wide range of cTnI detection (10 ag mL−1–100 ng mL−1) in human serum with a regression coefficient of 0.9716, rapid detection (5 min of binding time), and stable and highly reproducible bioelectrode response with RSD

Details

ISSN :
14365073 and 00263672
Volume :
188
Database :
OpenAIRE
Journal :
Microchimica Acta
Accession number :
edsair.doi...........6858579ed528f114bbdcfe00acd15c6f
Full Text :
https://doi.org/10.1007/s00604-021-04922-x