Back to Search Start Over

Fabrication of High-Resolution Gamma-Ray Metallic Magnetic Calorimeters with Ag:Er Sensor and Thick Electroplated Absorbers

Authors :
S. T. P. Boyd
J. A. Hall
G. B. Kim
Robin Cantor
Ruslan Hummatov
Stephan Friedrich
Source :
Journal of Low Temperature Physics. 193:752-757
Publication Year :
2018
Publisher :
Springer Science and Business Media LLC, 2018.

Abstract

We are developing metallic magnetic calorimeters for high-resolution gamma-ray spectroscopy for non-destructive assay of nuclear materials. Absorbers for these higher-energy photons can require substantial thickness to achieve adequate stopping power. We developed a new absorber fabrication process using dry-film photoresists to electroform cantilevered, thick absorbers. Gamma detectors with these absorbers have an energy resolution of 38 eV FWHM at 60 keV. In this report, we summarize modifications to STARCryo’s “Delta 1000” process for our devices and describe the new absorber fabrication process.

Details

ISSN :
15737357 and 00222291
Volume :
193
Database :
OpenAIRE
Journal :
Journal of Low Temperature Physics
Accession number :
edsair.doi...........5a87185497f3b81ab60ec4455d7df074
Full Text :
https://doi.org/10.1007/s10909-018-1946-8