Back to Search
Start Over
Geotech Smart24 data acquisition system input terminated noise seismic response adjusted test : StreckeisenSTS2-low and high gain, Guralp CMG3T and Geotech GS13 seismometers
- Publication Year :
- 2008
- Publisher :
- Office of Scientific and Technical Information (OSTI), 2008.
-
Abstract
- Sandia National Laboratories has tested, evaluated and reported on the Geotech Smart24 data acquisition system with active Fortezza crypto card data signing and authentication in SAND2008-. One test, Input Terminated Noise, allows us to characterize the self-noise of the Smart24 system. By computing the power spectral density (PSD) of the input terminated noise time series data set and correcting for the instrument response of different seismometers, the resulting spectrum can be compared to the USGS new low noise model (NLNM) of Peterson (1996), and determine the ability of the matched system of seismometer and Smart24 to be quiet enough for any general deployment location. Four seismometer models were evaluated: the Streckeisen STS2-Low and High Gain, Guralp CMG3T and Geotech GS13 models. Each has a unique pass-band as defined by the frequency band of the instrument corrected noise spectrum that falls below the new low-noise model.
Details
- Database :
- OpenAIRE
- Accession number :
- edsair.doi...........563f6e3d6a5ab0eecd30cec1b1027691