Back to Search
Start Over
Different material properties of cancellous bone influence analysis of glenoid component loosening: A finite element study
- Source :
- Malaysian Journal of Fundamental and Applied Sciences. 13:483-488
- Publication Year :
- 2017
- Publisher :
- Penerbit UTM Press, 2017.
-
Abstract
- Glenoid component loosening is one of common complications after total shoulder arthroplasty. In investigating the glenoid component loosening, the finite element study is one of methods that have been utilised by experts. Therefore, assigning material properties for all finite element models become crucial to avoid any misinterpretation which, later, lead to wrong prediction on the performance of glenoid implant. This study was conducted to achieve two objectives; (1) to analyse the effect of different bone properties towards micromotion and stress at implant and cement, and (2) to clarify simplification of bone properties in evaluating glenoid component loosening. A load of 750N was simulated at three different glenoid locations (centre – C, superior-anterior-SA, superior-posterior-SP) which imitate concentric and eccentric loadings for elderly people daily activities. Our result showed that large different in micromotion and stress at implant between orthotropic model and another two model (isotropic and full cortical) does not allow simplification for assigning material properties for bone. Thus, assigning cancellous bone as orthotropic material was a realistic material property to represent the real bone condition in evaluating glenoid implant loosening.
- Subjects :
- Orthodontics
Computer science
General Mathematics
medicine.medical_treatment
0206 medical engineering
030232 urology & nephrology
General Physics and Astronomy
02 engineering and technology
General Chemistry
Orthotropic material
020601 biomedical engineering
Arthroplasty
General Biochemistry, Genetics and Molecular Biology
Finite element method
Finite element study
Stress (mechanics)
03 medical and health sciences
0302 clinical medicine
medicine.anatomical_structure
medicine
Implant
General Agricultural and Biological Sciences
Material properties
Cancellous bone
Subjects
Details
- ISSN :
- 2289599X and 22895981
- Volume :
- 13
- Database :
- OpenAIRE
- Journal :
- Malaysian Journal of Fundamental and Applied Sciences
- Accession number :
- edsair.doi...........55cd9922419e469460d81ddd904d3b77