Back to Search Start Over

Sensitive Bolometric Detection of Atomic Hydrogen Fluxes

Authors :
E. J. Wolters
J. T. M. Walraven
Joris J. Berkhout
O. H. Höpfner
Source :
Japanese Journal of Applied Physics. 26:231
Publication Year :
1987
Publisher :
IOP Publishing, 1987.

Abstract

We developed a sensitive device to detect atomic hydrogen fluxes. The device is based on a measurement of the heating due to forced recombination of the atoms in the flux. We found that 60±10 % of the recombination heat is collected by the detector. The DC noise equivalent (1σ) flux Φ is 6×1010 at/s at a bandwidth of 0.4 Hz and a dynamic range of 1000 (Φmax=6×1013 at/s). The device may be used at ambient temperatures below 450 mK. We discuss possible improvements of the design.

Details

ISSN :
13474065 and 00214922
Volume :
26
Database :
OpenAIRE
Journal :
Japanese Journal of Applied Physics
Accession number :
edsair.doi...........55727a275f6b45f9ed0923619d2e27d6
Full Text :
https://doi.org/10.7567/jjaps.26s3.231