Back to Search Start Over

A simplified computer program for HPGe detector absolute peak efficiency calibration

Authors :
Lih-Jen Kang
Tien-Ko Wang
Tai-Yin Chen
Source :
Radiation Physics and Chemistry. 42:1039-1042
Publication Year :
1993
Publisher :
Elsevier BV, 1993.

Abstract

A computer program ESOLAN has been developed for HPGe detector peak-efficiency calibration in various counting geometries. The program was written for use in personal computers(PCs). Experiments designate that the efficiency calibration, with the aid of this program, can be constructed with an overall uncertainty of about 3% or lower for gamma rays ranging from 80 KeV to 3 MeV.

Details

ISSN :
0969806X
Volume :
42
Database :
OpenAIRE
Journal :
Radiation Physics and Chemistry
Accession number :
edsair.doi...........5056a6ae3077b642afaffda4b38adbf1
Full Text :
https://doi.org/10.1016/0969-806x(93)90429-x